New ZEISS CIRRUS photo One system for fundus imaging and OCT
[vc_row][vc_column width=”1/1″][vc_column_text]
ZEISS CIRRUS Photo One System for Fundus Imaging and OCT
[/vc_column_text][vc_tabs][vc_tab title=”Highlights” tab_id=”1450206237-1-7″][vc_column_text]
One System for Fundus Imaging and OCT
Broader clinical insights, greater diagnostic certainty and added practice value – the new CIRRUS™ photo
from ZEISS delivers all that in a single, integrated system for both fundus imaging and OCT. CIRRUS photo combines a full mydriatic/non-mydriatic fundus camera with proven CIRRUS HD-OCT technology in one compact and highly versatile system.
Visualize findings from various modalities. Correlate data from high-density OCT cubes, thickness and layer maps with results from superb color fundus images as well as fundus autofluorescence and fluorescein angiography* images. All in one convenient sitting.
Have it all in a single system for fundus imaging and OCT.
[/vc_column_text][/vc_tab][vc_tab title=”Specifications” tab_id=”1450206237-2-51″][vc_column_text]
Main System
Field angle | 45° and 30° |
Pupil diameter | ≥ 4.0 mm; ≥ 3.3 mm (30° small pupil mode) ≥ 2.0 mm for OCT scans only |
Refractive error compensation | +35 D … -35 D, continuous |
Working distance | 40 mm (patient’s eye – front lens) |
Fixation targets | External and internal |
Internal | Attention mode and free position or programmed sequences |
Database | Patient information and images with field angle, FA time, R/L recognition and date of visit are stored |
Monitor | 23” TFT (1920 x 1200) |
Instrument table | Asymmetric, suitable for wheelchairs |
Accessories | Network printer, sliding keyboard shelf, network isolator, FORUM eye care data management system |
Fundus Camera
Capture modes | Color, red-free, blue, red and fundus autofluorescence pictures, as well as pictures of the anterior segment, CIRRUS photo 800 only: + fluorescein angiography and ICG angiography |
Filters | Filters for green, blue and fundus autofluorescence images, UV/IR barrier filters CIRRUS photo 800 only: + FA + ICGA: exciter and barrier filters |
Capture sequence | From 1.5 seconds (depends on flash energy) |
Capture sensor | CCD 5.0 megapixels |
Xenon flash lamp | 16 flash levels (30 Ws) CIRRUS photo 800 only: 24 flash levels (80 Ws) |
OCT
Technology | Spectral domain OCT |
Optical source | Superluminescent diode (SLD), 840 nm |
Scan speed | 27,000 A-scan per second |
A-scan depth | 2.0 mm (in tissue), 1024 points |
Resolution | Axial 5 μm (in tissue), transverse 15 μm (in tissue) |
Computer
Operating system | Windows Embedded Standard 7 |
Hard drive | Storage of over 30,000 fundus images with OCT cube scans (present size: 320 GB) |
Interfaces | USB ports and network connectors, DVI port |
Export/import | Image formats: BMP, TIFF, JPEG, PNG Patient list, DICOM MWL, DICOM storage |
Dimensions
Main unit | 410 mm x 480 mm x 680 mm (W 16.1 x D 18.9 x H 26.8 inches) |
Weight (main unit) | 33 kg (72.7 lbs) |
Rated voltage | 100 … 240 V ±10% |
Frequency | 50 / 60 Hz |
Power consumption | 400 VA (w/o instrument table) |